HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY ANALYTICAL METHOD VALIDATION FOR GLUTARALDEHYDE AND BENZALKONIUM CHLORIDE IN DISINFECTANTS

Authors

  • Baitha Palanggatan Maggadani Department of Pharmacy, Faculty of Pharmacy, Universitas Indonesia, Depok, 16424, Indonesia
  • Harmita . Department of Pharmacy, Faculty of Pharmacy, Universitas Indonesia, Depok, 16424, Indonesia
  • Maizura Isfadhila Department of Pharmacy, Faculty of Pharmacy, Universitas Indonesia, Depok, 16424, Indonesia

DOI:

https://doi.org/10.22159/ijap.2018.v10s1.55

Keywords:

Glutaraldehyde, Benzalkonium chloride, High-performance liquid chromatography, Optimization, Validation, Disinfectant

Abstract

Objective: The aim of this study was to produce a selective, accurate, and faster high-performance liquid chromatography (HPLC) analytical method
for benzalkonium chloride and glutaraldehyde in disinfectants using ultraviolet (UV)-visible detection.
Methods: Glutaraldehyde has no chromophore, so it was first derivatized using 2,4 dinitro phenylhydrazine. Acetonitrile:water (75:25) was used as
the mobile phase for glutaraldehyde and acetonitrile-acetate pH 4 (75:25) for benzalkonium chloride, both at a flow rate of 1.2 mL/min. The optimized
assay was validated with respect to accuracy, precision, linearity, selectivity, limit of quantitation (LOQ), and limit of detection (LOD).
Results: The method was linear for benzalkonium chloride, with correlation coefficient of 0.9995, LOD of 14.55 ppm, and LOQ of 48.51 ppm. The
correlation coefficient for glutaraldehyde was 0.9995, with LOD of 0.49 ppm and LOQ of 1.64 ppm. Accuracy was between 98% and 102%, and
precision was below 2% for both the tests.
Conclusion: The HPLC analytical method for benzalkonium chloride and glutaraldehyde in disinfectants using UV-visible detection in this research
was successful to produce a selective, accurate, and faster method.

Downloads

Download data is not yet available.

References

Sweetman, S. Martindale. London: PhP Pharmaceutical Press; 2009.

Koyama K, Shimazu T. Benzalkonium Chloride. Berlin Heidelberg:

Springe-Verlag; 2005. p. 407-13.

Directorate General of Livestock Product Development Indonesian

Ministry of Agriculture. Indonesian Veterinary Drug Pharmacopoeia.

Jakarta: Indonesian Ministry of Agriculture; 2008.

Labranche LP, Dumont SN, Levesque S, Carrier A. Rapid determination

of total benzalkonium chloride content in ophthalmic formulation.

J Pharm Biomed Anal 2007;43:989-93.

Menet MC, Gueylard D, Fievet MH, Thuillier A. Fast specific separation

and sensitive quantification of bactericidal and sporicidal aldehydes by

high-performance liquid chromatography: Example of glutaraldehyde

determination. J Chromatogr Sep Tech 1997;692:79-86.

Harmita. Physicochemistry Analysis. Depok: Faculty of Pharmacy,

Universitas Indonesia; 2012.

Rediatning W, Kartini N. Analysis of amino acids with high performance

liquid chromatography derivatized prakolom and pascakolom. Proc

ITB 1987;20:1-19.

Rao N, Gawde KD. Method development and force degradation

studies for simultaneous estimation of salbutamol sulfate, etofylline

and bromhexine hydrochloride in pharmaceutical dosage form using

reversed-phase high-performance liquid chromatography method.

Asian J Pharm Clin Res 2018;11:378-82.

Shaikh S, Jain V. A novel reverse-phase high-performance liquid

chromatographic method for simultaneous estimation of ellagic acid,

quercetin, and piperine in ayurvedic formulations. Asian J Pharm Clin

Res 2018;11:312-7.

Ramesha B, Reddy KR, Unni KM, Ankolekar V, Anuradha P,

Amith KM. A validated UHPLC method for the determination of

atorvastatin acetonide tert-butyl ester and 4-fluoro-alpha-(2-methyl-1-

oxopropyl)-gamma-oxo-n, beta-diphenylbenzene butaneamide. Asian J

Pharm Clin Res 2012;5:115-22.

ICH. ICH Topic Q2 (R1) Validation of Analytical Procedures: Text and

Methodology. International Conference on Harmonization, 1994; 1996.

Thompson M, Ellison SL, Wood R. Harmonized guidelines for singlelaboratory

validation of methods of analysis (IUPAC Technical Report).

Pure Appl Chem 2002;74:835-55.

Bretnall AE, Clarke GS. Validation of analytical test methods. In: Ahuja

S, Scypinski S, editors. Handbook of Modern Pharmaceutical Analysis,

Separation Science and Technology Series. 2nd ed. Waltham, MA, USA:

Academic Press (Elsevier); 2011. p. 9-457.

Anderson J, Berthod A, Pine V, Stalcup AM. Analytical Separation

Science. Harlow, UK: Prentice-Hall; 2016.

Published

20-12-2018

How to Cite

Maggadani, B. P., ., H., & Isfadhila, M. (2018). HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY ANALYTICAL METHOD VALIDATION FOR GLUTARALDEHYDE AND BENZALKONIUM CHLORIDE IN DISINFECTANTS. International Journal of Applied Pharmaceutics, 10(1), 248–251. https://doi.org/10.22159/ijap.2018.v10s1.55

Issue

Section

Original Article(s)